Description
The Infant Incubator YR06213 is a state-of-the-art medical device designed to provide a safe and controlled environment for newborns. Equipped with a 7-inch LCD touch screen control panel, this incubator allows precise monitoring and adjustment of temperature, ensuring the utmost safety and comfort for infants. Its air and skin temperature are servo-controlled by a computer, offering unparalleled accuracy. The incubator includes features like self-check alarms, a removable humidity trough easy for cleaning, and a transfusion shelf with an added guardrail for convenience and safety. Additionally, the RS-232 connector, humidity display, and optional phototherapy make this an indispensable tool for neonatal care.
Market Price
In the competitive market of infant care technology, prices for devices like the Infant Incubator YR06213 generally range from $10,500 to $11,400. These incubators are pivotal in newborn care, offering advanced features and reliability that justify their price range. For accurate pricing and to explore available discounts, it’s recommended to request a quote through our platform, Kalstein Plus.
Frequently Asked Questions
Q: What power supply does the Infant Incubator YR06213 require?
A: The incubator operates on an electrical requirement of ~220V 50Hz.
Q: Can the incubator’s mattress be adjusted?
A: Yes, it features a free step mattress tilting adjustment for added comfort and positioning.
Q: Is there an alarm function?
A: Yes, it includes various alarms, such as air and skin temperature alarms, to ensure infant safety.
Advantages and Disadvantages
The Infant Incubator YR06213 offers numerous advantages, including precise temperature control, ease of cleaning, and optional advanced features like phototherapy. However, its premium technology may come with a higher price point compared to more basic models.
Product Use in the Field
This incubator is widely used in neonatal intensive care units (NICUs) across the globe. It provides stable and controlled environmental conditions crucial for the development and treatment of premature and vulnerable infants.
Recommendations
To maintain the optimal functionality of the Infant Incubator YR06213, regular maintenance checks and calibration should be performed. Proper training for the medical staff is also essential to ensure the effective and safe use of the device. Consider leveraging the automated quote system via Kalstein Plus for easy procurement.
Features
- 7-inch LCD touch screen control panel
- Air temperature and skin temperature servo-controlled by computer
- Self-check alarms
- Removable humidity trough
- Optional phototherapy
- RS-232 connector
- Humidity display and control
- Transfusion shelf and guardrail included
- Multiple adjustable elements for infant comfort
- Optional accessories: drawer, cabinet, electric weighing function, and electrical lift
Technical Specifications
Model | YR06213 |
Electrical requirement | ~220V 50Hz |
Heater Power | 550VA |
Air mode temperature range | 25.0°C-37.0°C |
Skin mode temperature range | 34.0°C-37.0°C |
Air mode temperature alarm | <38.0°C |
Skin mode temperature alarm | < 40.0°C(Air temperature) |
Air mode high temperature range | 37.1°C~39.0°C |
Skin mode high temperature range | 37.1°C~38.0°C |
Hight temperature mode alarm | < 40.0°C (Air temperature) |
Air mode temperature display range | 20.0°C-45.0°C |
Skin mode temperature display range | 25.0°C~45.0°C |
Air mode alarm | ±3.0°C |
Skin mode alarm | ±1.0°C |
Skin temperature sensor accuracy | ±0.3°C |
Temperature control accuracy | ≤0.5°C |
Mattress temperature uniformity | ≤0.8°C |
Humidity control range | 30-90%RH |
Humidity display range | 0-99.9%RH |